Shortening of paced QRS duration after electrocardiographic optimization of left ventricular pacing vector in patients treated with Cardiac Resynchronization Therapy

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

Background: Choice of left ventricular pacing vector (LVPV) affects the QRS-duration (QRSd) in patients with Cardiac Resynchronization Therapy (CRT). It is not known whether testing all LVPVs reduces QRSd compared to device-based “standard-programming”.

Methods: In patients implanted with CRT several ECGs were recorded for each usable LVPV (no phrenic nerve stimulation and threshold <3.5 V) and during “standard-programming” after device-based optimization of AV/VV delays.

Results: 22 consecutive patients were included. Average QRSd reduction after CRT + “standard-programming” was 27.3 ± 22 ms. Additional QRSd-reduction was possible in 4 patients by changing the LVPV, and in 5 other patients after optimization of AV- and VV delays without changing LVPV.

Conclusions: Shortening of QRSd compared to “standard-programming” was possible approximately 40% of these patients treated with CRT by testing all LVPVs and re-optimizing AV/VV delays during follow-up. Studies of clinical effects are needed.

OriginalsprogEngelsk
TidsskriftJournal of Electrocardiology
Vol/bind51
Udgave nummer4
Sider (fra-til)628-633
Antal sider6
ISSN0022-0736
DOI
StatusUdgivet - 2018

ID: 216310029