Removal of Polysorbate 80 by complexation prior to LC-MS analysis

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

The presence of Polysorbate 80 in samples can challenge liquid chromatography-electrospray ionisation mass spectrometry (LC-ESI-MS) analysis as it is easily ionised and detected. In this study, we demonstrate that interference from Polysorbate 80 can be reduced by complexation with a metal ion followed by precipitation by thiocyanate. The precipitation procedure was tested on a mixture of low molecular weight compounds (e.g. amino acids and non-amino organic acids) and it was shown that none of the tested compounds were precipitated.

OriginalsprogEngelsk
TidsskriftAnalusis - European Journal of Analytical Chemistry
Vol/bind408
Udgave nummer9
Sider (fra-til)2303-2307
Antal sider5
ISSN0365-4877
DOI
StatusUdgivet - 2016

ID: 169136315