Effect of make-up gas in on-column atomic emission spectrometric detection for capillary gas chromatography

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

The effect of make-up gas was studied for capillary gas chromatography (GC) coupled with on-column atomic emission spectrometric detection (AED), where a 350 kHz plasma was sustained inside the end of a 0.32 mm i.d. capillary GC column. For selective detection of Cl and Br both the plasma background characteristics and elemental responses were stabilized during temperature programmed analysis when 10-15 ml min-1 of helium were added as a make-up gas. In addition, Chromatographic peaks were protected against tailing, while Cl and Br were detected at the 4 pg s-1 level.

OriginalsprogEngelsk
TidsskriftJournal of Analytical Atomic Spectrometry
Vol/bind11
Udgave nummer2
Sider (fra-til)117-121
Antal sider5
ISSN0267-9477
DOI
StatusUdgivet - feb. 1996

ID: 231656434