Diagnostic accuracy of semi-automatic quantitative metrics as an alternative to expert reading of CT myocardial perfusion in the CORE320 study

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

  • Mohammad R Ostovaneh
  • Andrea L Vavere
  • Vishal C Mehra
  • Kofoed, Klaus Fuglsang
  • Matthew B Matheson
  • Armin Arbab-Zadeh
  • Yasuko Fujisawa
  • Joanne D Schuijf
  • Carlos E Rochitte
  • Arthur J Scholte
  • Kakuya Kitagawa
  • Marc Dewey
  • Christopher Cox
  • Marcelo F DiCarli
  • Richard T George
  • Joao A C Lima

AIMS: To determine the diagnostic accuracy of semi-automatic quantitative metrics compared to expert reading for interpretation of computed tomography perfusion (CTP) imaging.

METHODS: The CORE320 multicenter diagnostic accuracy clinical study enrolled patients between 45 and 85 years of age who were clinically referred for invasive coronary angiography (ICA). Computed tomography angiography (CTA), CTP, single photon emission computed tomography (SPECT), and ICA images were interpreted manually in blinded core laboratories by two experienced readers. Additionally, eight quantitative CTP metrics as continuous values were computed semi-automatically from myocardial and blood attenuation and were combined using logistic regression to derive a final quantitative CTP metric score. For the reference standard, hemodynamically significant coronary artery disease (CAD) was defined as a quantitative ICA stenosis of 50% or greater and a corresponding perfusion defect by SPECT. Diagnostic accuracy was determined by area under the receiver operating characteristic curve (AUC).

RESULTS: Of the total 377 included patients, 66% were male, median age was 62 (IQR: 56, 68) years, and 27% had prior myocardial infarction. In patient based analysis, the AUC (95% CI) for combined CTA-CTP expert reading and combined CTA-CTP semi-automatic quantitative metrics was 0.87(0.84-0.91) and 0.86 (0.83-0.9), respectively. In vessel based analyses the AUC's were 0.85 (0.82-0.88) and 0.84 (0.81-0.87), respectively. No significant difference in AUC was found between combined CTA-CTP expert reading and CTA-CTP semi-automatic quantitative metrics in patient based or vessel based analyses(p > 0.05 for all).

CONCLUSION: Combined CTA-CTP semi-automatic quantitative metrics is as accurate as CTA-CTP expert reading to detect hemodynamically significant CAD.

OriginalsprogEngelsk
TidsskriftJournal of Cardiovascular Computed Tomography
Vol/bind12
Udgave nummer3
Sider (fra-til)212-219
Antal sider8
ISSN1934-5925
DOI
StatusUdgivet - 2018

ID: 216201840