Velocity estimation of the main portal vein with Transverse Oscillation

Research output: Chapter in Book/Report/Conference proceedingArticle in proceedingsResearchpeer-review

Standard

Velocity estimation of the main portal vein with Transverse Oscillation. / Brandt, Andreas Hjelm; Hansen, Kristoffer Lindskov; Nielsen, Michael Bachmann; Jensen, Jorgen Arendt.

2015 IEEE International Ultrasonics Symposium, IUS 2015. Institute of Electrical and Electronics Engineers Inc., 2015. 7329104 (2015 IEEE International Ultrasonics Symposium, IUS 2015).

Research output: Chapter in Book/Report/Conference proceedingArticle in proceedingsResearchpeer-review

Harvard

Brandt, AH, Hansen, KL, Nielsen, MB & Jensen, JA 2015, Velocity estimation of the main portal vein with Transverse Oscillation. in 2015 IEEE International Ultrasonics Symposium, IUS 2015., 7329104, Institute of Electrical and Electronics Engineers Inc., 2015 IEEE International Ultrasonics Symposium, IUS 2015, IEEE International Ultrasonics Symposium, IUS 2015, Taipei, Taiwan, Province of China, 21/10/2015. https://doi.org/10.1109/ULTSYM.2015.0068

APA

Brandt, A. H., Hansen, K. L., Nielsen, M. B., & Jensen, J. A. (2015). Velocity estimation of the main portal vein with Transverse Oscillation. In 2015 IEEE International Ultrasonics Symposium, IUS 2015 [7329104] Institute of Electrical and Electronics Engineers Inc.. 2015 IEEE International Ultrasonics Symposium, IUS 2015 https://doi.org/10.1109/ULTSYM.2015.0068

Vancouver

Brandt AH, Hansen KL, Nielsen MB, Jensen JA. Velocity estimation of the main portal vein with Transverse Oscillation. In 2015 IEEE International Ultrasonics Symposium, IUS 2015. Institute of Electrical and Electronics Engineers Inc. 2015. 7329104. (2015 IEEE International Ultrasonics Symposium, IUS 2015). https://doi.org/10.1109/ULTSYM.2015.0068

Author

Brandt, Andreas Hjelm ; Hansen, Kristoffer Lindskov ; Nielsen, Michael Bachmann ; Jensen, Jorgen Arendt. / Velocity estimation of the main portal vein with Transverse Oscillation. 2015 IEEE International Ultrasonics Symposium, IUS 2015. Institute of Electrical and Electronics Engineers Inc., 2015. (2015 IEEE International Ultrasonics Symposium, IUS 2015).

Bibtex

@inproceedings{d1be5dcbbb194d26a8caca807b50ba7c,
title = "Velocity estimation of the main portal vein with Transverse Oscillation",
abstract = "This study evaluates if Transverse Oscillation (TO) can provide reliable and accurate peak velocity estimates of blood flow the main portal vein. TO was evaluated against the recommended and most widely used technique for portal flow estimation, Spectral Doppler Ultrasound (SDU). The main portal vein delivers blood from the bowls to the liver, and patients with certain liver diseases have decreased flow in the portal vein. Errors in velocity estimation with SDU are well described, when the beam-to-flow angle is >70 degrees. TO estimates the flow angle independently and is not limited by the beam-to-flow angle. It is less operators depended, as no angle correction is necessary. TO measurements were performed with a 3 MHz convex probe (BK medical 8820e, Herlev, Denmark) connected to the experimental ultrasound scanner SARUS (Synthetic Aperture Real-time Ultrasound Scanner). SDU velocity measurements were performed with a commercial ultrasound scanner (BK 3000, BK Ultrasound, Herlev Denmark) and a convex probe (BK ultrasound 6C2, Herlev, Denmark). Ten healthy volunteers were scanned, and recordings of the portal flow during 3-5 heartbeats were conducted with an intercostal and subcostal view. Intercostal TO peak velocities were not significantly different from SDU peak velocities (TO=0.203m/s, SDU=0.202m/s, p=0.94). Subcostal and Intercostal obtained TO values were not significantly different (intercostal mean TO=0.203m/s, subcostal mean TO=0.180m/s, p=0.26). SDU values obtained intercostal and subcostal were significantly different (intercostal mean SDU=0.202m/s, subcostal mean SDU=0.320m/s, p<0.001). Standard deviation for TO beam-to-flow angle was 10.3°-91.5°, indicating a large beam-to-flow angle variability in the portal vein. This can affect the peak velocity estimation, and is not addressed in SDU. The TO convex array implementation provides the first vector velocity measurements below 60mm (mean 89mm), and is a useful alternative for flow estimation in abdominal ultrasound. It may provide new information of abdominal fluid dynamics and yield both velocity and angle estimates for a more realistic flow characterization.",
keywords = "Main Portal Vein, Spectral Doppler, Transverve Oscillation, Ultrasound Velocity Estimation, Vector Velocity Estimation",
author = "Brandt, {Andreas Hjelm} and Hansen, {Kristoffer Lindskov} and Nielsen, {Michael Bachmann} and Jensen, {Jorgen Arendt}",
note = "Publisher Copyright: {\textcopyright} 2015 IEEE.; IEEE International Ultrasonics Symposium, IUS 2015 ; Conference date: 21-10-2015 Through 24-10-2015",
year = "2015",
month = nov,
day = "13",
doi = "10.1109/ULTSYM.2015.0068",
language = "English",
series = "2015 IEEE International Ultrasonics Symposium, IUS 2015",
booktitle = "2015 IEEE International Ultrasonics Symposium, IUS 2015",
publisher = "Institute of Electrical and Electronics Engineers Inc.",

}

RIS

TY - GEN

T1 - Velocity estimation of the main portal vein with Transverse Oscillation

AU - Brandt, Andreas Hjelm

AU - Hansen, Kristoffer Lindskov

AU - Nielsen, Michael Bachmann

AU - Jensen, Jorgen Arendt

N1 - Publisher Copyright: © 2015 IEEE.

PY - 2015/11/13

Y1 - 2015/11/13

N2 - This study evaluates if Transverse Oscillation (TO) can provide reliable and accurate peak velocity estimates of blood flow the main portal vein. TO was evaluated against the recommended and most widely used technique for portal flow estimation, Spectral Doppler Ultrasound (SDU). The main portal vein delivers blood from the bowls to the liver, and patients with certain liver diseases have decreased flow in the portal vein. Errors in velocity estimation with SDU are well described, when the beam-to-flow angle is >70 degrees. TO estimates the flow angle independently and is not limited by the beam-to-flow angle. It is less operators depended, as no angle correction is necessary. TO measurements were performed with a 3 MHz convex probe (BK medical 8820e, Herlev, Denmark) connected to the experimental ultrasound scanner SARUS (Synthetic Aperture Real-time Ultrasound Scanner). SDU velocity measurements were performed with a commercial ultrasound scanner (BK 3000, BK Ultrasound, Herlev Denmark) and a convex probe (BK ultrasound 6C2, Herlev, Denmark). Ten healthy volunteers were scanned, and recordings of the portal flow during 3-5 heartbeats were conducted with an intercostal and subcostal view. Intercostal TO peak velocities were not significantly different from SDU peak velocities (TO=0.203m/s, SDU=0.202m/s, p=0.94). Subcostal and Intercostal obtained TO values were not significantly different (intercostal mean TO=0.203m/s, subcostal mean TO=0.180m/s, p=0.26). SDU values obtained intercostal and subcostal were significantly different (intercostal mean SDU=0.202m/s, subcostal mean SDU=0.320m/s, p<0.001). Standard deviation for TO beam-to-flow angle was 10.3°-91.5°, indicating a large beam-to-flow angle variability in the portal vein. This can affect the peak velocity estimation, and is not addressed in SDU. The TO convex array implementation provides the first vector velocity measurements below 60mm (mean 89mm), and is a useful alternative for flow estimation in abdominal ultrasound. It may provide new information of abdominal fluid dynamics and yield both velocity and angle estimates for a more realistic flow characterization.

AB - This study evaluates if Transverse Oscillation (TO) can provide reliable and accurate peak velocity estimates of blood flow the main portal vein. TO was evaluated against the recommended and most widely used technique for portal flow estimation, Spectral Doppler Ultrasound (SDU). The main portal vein delivers blood from the bowls to the liver, and patients with certain liver diseases have decreased flow in the portal vein. Errors in velocity estimation with SDU are well described, when the beam-to-flow angle is >70 degrees. TO estimates the flow angle independently and is not limited by the beam-to-flow angle. It is less operators depended, as no angle correction is necessary. TO measurements were performed with a 3 MHz convex probe (BK medical 8820e, Herlev, Denmark) connected to the experimental ultrasound scanner SARUS (Synthetic Aperture Real-time Ultrasound Scanner). SDU velocity measurements were performed with a commercial ultrasound scanner (BK 3000, BK Ultrasound, Herlev Denmark) and a convex probe (BK ultrasound 6C2, Herlev, Denmark). Ten healthy volunteers were scanned, and recordings of the portal flow during 3-5 heartbeats were conducted with an intercostal and subcostal view. Intercostal TO peak velocities were not significantly different from SDU peak velocities (TO=0.203m/s, SDU=0.202m/s, p=0.94). Subcostal and Intercostal obtained TO values were not significantly different (intercostal mean TO=0.203m/s, subcostal mean TO=0.180m/s, p=0.26). SDU values obtained intercostal and subcostal were significantly different (intercostal mean SDU=0.202m/s, subcostal mean SDU=0.320m/s, p<0.001). Standard deviation for TO beam-to-flow angle was 10.3°-91.5°, indicating a large beam-to-flow angle variability in the portal vein. This can affect the peak velocity estimation, and is not addressed in SDU. The TO convex array implementation provides the first vector velocity measurements below 60mm (mean 89mm), and is a useful alternative for flow estimation in abdominal ultrasound. It may provide new information of abdominal fluid dynamics and yield both velocity and angle estimates for a more realistic flow characterization.

KW - Main Portal Vein

KW - Spectral Doppler

KW - Transverve Oscillation

KW - Ultrasound Velocity Estimation

KW - Vector Velocity Estimation

UR - http://www.scopus.com/inward/record.url?scp=84962030433&partnerID=8YFLogxK

U2 - 10.1109/ULTSYM.2015.0068

DO - 10.1109/ULTSYM.2015.0068

M3 - Article in proceedings

AN - SCOPUS:84962030433

T3 - 2015 IEEE International Ultrasonics Symposium, IUS 2015

BT - 2015 IEEE International Ultrasonics Symposium, IUS 2015

PB - Institute of Electrical and Electronics Engineers Inc.

T2 - IEEE International Ultrasonics Symposium, IUS 2015

Y2 - 21 October 2015 through 24 October 2015

ER -

ID: 331499363