Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature-controlled irrigated tip ablation, power-controlled irrigated tip ablation, and standard temperature-controlled ablation
Publikation: Bidrag til tidsskrift › Tidsskriftartikel › Forskning › fagfællebedømt
Standard
Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature-controlled irrigated tip ablation, power-controlled irrigated tip ablation, and standard temperature-controlled ablation. / Petersen, H H; Chen, X; Pietersen, A; Svendsen, Jesper Hastrup; Haunsø, S.
I: Pacing and Clinical Electrophysiology, Bind 23, Nr. 1, 01.01.2000, s. 8-17.Publikation: Bidrag til tidsskrift › Tidsskriftartikel › Forskning › fagfællebedømt
Harvard
APA
Vancouver
Author
Bibtex
}
RIS
TY - JOUR
T1 - Tissue temperatures and lesion size during irrigated tip catheter radiofrequency ablation: an in vitro comparison of temperature-controlled irrigated tip ablation, power-controlled irrigated tip ablation, and standard temperature-controlled ablation
AU - Petersen, H H
AU - Chen, X
AU - Pietersen, A
AU - Svendsen, Jesper Hastrup
AU - Haunsø, S
PY - 2000/1/1
Y1 - 2000/1/1
N2 - The limited success rate of radiofrequency catheter ablation in patients with ventricular tachycardias related to structural heart disease may be increased by enlarging the lesion size. Irrigated tip catheter ablation is a new method for enlarging the size of the lesion. It was introduced in the power-controlled mode with high power and high infusion rate, and is associated with an increased risk of crater formation, which is related to high tissue temperatures. The present study explored the tissue temperatures during temperature-controlled irrigated tip ablation, comparing it with standard temperature-controlled ablation and power-controlled irrigated tip ablation. In vitro strips of porcine left ventricular myocardium were ablated. Temperature-controlled irrigated tip ablation at target temperatures 60 degrees C, 70 degrees C, and 80 degrees C with infusion of 1 mL saline/min were compared with standard temperature-controlled ablation at 70 degrees C and power-controlled irrigated tip ablation at 40 W, and infusion of 20 mL/min. Lesion size and tissue temperatures were significantly higher during all modes of irrigated tip ablation compared with standard temperature-controlled ablation (P <0.05). Lesion volume correlated positively with tissue temperature (r = 0.87). The maximum recorded tissue temperature was always 1 mm from the ablation electrode and was 67 +/- 4 degrees C for standard ablation and 93 +/- 6 degrees C, 99 +/- 6 degrees C, and 115 +/- 13 degrees C for temperature-controlled irrigated tip ablation at 60 degrees C, 70 degrees C, and 80 degrees C, respectively, and 112 +/- 12 degrees C for power-controlled irrigated tip ablation, which for irrigated tip ablation was significantly higher than tip temperature (P <0.0001). Crater formation only occurred at tissue temperatures > 100 degrees C. We conclude that irrigated tip catheter ablation increases lesion size and tissue temperatures compared with standard ablation in the temperature-controlled mode at the same or higher target temperatures and in the power-controlled mode. Furthermore, tissue temperature and delivered power are the best indicators of lesion volume during temperature-controlled ablation.
AB - The limited success rate of radiofrequency catheter ablation in patients with ventricular tachycardias related to structural heart disease may be increased by enlarging the lesion size. Irrigated tip catheter ablation is a new method for enlarging the size of the lesion. It was introduced in the power-controlled mode with high power and high infusion rate, and is associated with an increased risk of crater formation, which is related to high tissue temperatures. The present study explored the tissue temperatures during temperature-controlled irrigated tip ablation, comparing it with standard temperature-controlled ablation and power-controlled irrigated tip ablation. In vitro strips of porcine left ventricular myocardium were ablated. Temperature-controlled irrigated tip ablation at target temperatures 60 degrees C, 70 degrees C, and 80 degrees C with infusion of 1 mL saline/min were compared with standard temperature-controlled ablation at 70 degrees C and power-controlled irrigated tip ablation at 40 W, and infusion of 20 mL/min. Lesion size and tissue temperatures were significantly higher during all modes of irrigated tip ablation compared with standard temperature-controlled ablation (P <0.05). Lesion volume correlated positively with tissue temperature (r = 0.87). The maximum recorded tissue temperature was always 1 mm from the ablation electrode and was 67 +/- 4 degrees C for standard ablation and 93 +/- 6 degrees C, 99 +/- 6 degrees C, and 115 +/- 13 degrees C for temperature-controlled irrigated tip ablation at 60 degrees C, 70 degrees C, and 80 degrees C, respectively, and 112 +/- 12 degrees C for power-controlled irrigated tip ablation, which for irrigated tip ablation was significantly higher than tip temperature (P <0.0001). Crater formation only occurred at tissue temperatures > 100 degrees C. We conclude that irrigated tip catheter ablation increases lesion size and tissue temperatures compared with standard ablation in the temperature-controlled mode at the same or higher target temperatures and in the power-controlled mode. Furthermore, tissue temperature and delivered power are the best indicators of lesion volume during temperature-controlled ablation.
KW - Animals
KW - Body Temperature
KW - Catheter Ablation
KW - Disease Models, Animal
KW - Heart Conduction System
KW - Heart Ventricles
KW - Swine
KW - Tachycardia, Ventricular
KW - Therapeutic Irrigation
KW - Treatment Outcome