Measurement of the arterial concentration of Gd-DTPA using MRI: a step toward quantitative perfusion imaging.

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

  • T Fritz-Hansen
  • Rostrup, Egill
  • H B Larsson
  • Lise Søndergaard
  • P Ring
  • O Henriksen
A noninvasive method using an inversion recovery turbo-FLASH for dynamic measurement of the arterial input function represented by the bolus passage of Gd-DTPA in the descending aorta is presented, and the results are compared with the input function obtained by arterial blood samples. A good accordance between the two input functions was found, indicating that it is possible to measure the input function to the myocardium using MRI. A variation between the two concentration curves of 5% at upslope, 2.7% at peak point, and < 7% at downslope was found. The study also indicates that a short inversion time < 250 ms has to be used to ensure correct measurement of peak concentration.
Bidragets oversatte titelMeasurement of the arterial concentration of Gd-DTPA using MRI: a step toward quantitative perfusion imaging.
OriginalsprogEngelsk
TidsskriftMagnetic Resonance in Medicine
Vol/bind36
Udgave nummer2
Sider (fra-til)225-231
Antal sider7
ISSN0740-3194
StatusUdgivet - 1996