99mTc-albumin can replace 125I-albumin to determine plasma volume repeatedly

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

OBJECTIVE: Plasma volume assessment may be of importance in several disorders. The purpose of the present study was to compare the reliability of plasma volume measurements by technetium-labeled human serum albumin ((99m)Tc-HSA) with a simultaneously performed plasma volume determination with iodine-labeled human serum albumin ((125)I-HSA).

MATERIALS AND METHODS: In 15 healthy volunteers, simultaneous plasma volume measurements with (99m)Tc-HSA and (125)I-HSA were performed after ½ hour in the supine position. Blood samples were obtained 10, 15, 20, and 30 minutes after the injection for accurate retropolation from the plasma counts to time zero to correct for leakage of the isotopes from the circulation.

RESULTS: The mean difference (bias) between plasma volume measured with (125)I-albumin and (99m)Tc-albumin was 8 ml (0.1 ml/kg) with limits of agreement (bias ±1.96 SD) ranging from -181-196 ml (-2.3-2.5 ml/kg). The tracer disappearance rate was significantly higher with (99m)Tc-albumin (-23.1±7.1%/h) than with (125)I-albumin (-6.7±3.6%/h) (p <0.001).

CONCLUSION: This study demonstrates that (99m)Tc-HSA can replace (125)I-HSA for single measurements of plasma volume in healthy volunteers. It needs to be emphasized however, that repeated blood sampling for 1/2 hour after injection of the tracer is required to correct for the disappearance of (99m)Tc and (99m)Tc-HSA from the circulation.

OriginalsprogEngelsk
TidsskriftScandinavian Journal of Clinical & Laboratory Investigation
Vol/bind72
Udgave nummer6
Sider (fra-til)447-51
Antal sider5
ISSN0036-5513
DOI
StatusUdgivet - okt. 2012

ID: 120835257