Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

Standard

Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol. / Olsson, Hampus; Novén, Mikael; Lätt, Jimmy; Wirestam, Ronnie; Helms, Gunther.

I: Tomography, Bind 7, Nr. 3, 2021, s. 434-451.

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

Harvard

Olsson, H, Novén, M, Lätt, J, Wirestam, R & Helms, G 2021, 'Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol', Tomography, bind 7, nr. 3, s. 434-451. https://doi.org/10.3390/tomography7030038

APA

Olsson, H., Novén, M., Lätt, J., Wirestam, R., & Helms, G. (2021). Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol. Tomography, 7(3), 434-451. https://doi.org/10.3390/tomography7030038

Vancouver

Olsson H, Novén M, Lätt J, Wirestam R, Helms G. Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol. Tomography. 2021;7(3):434-451. https://doi.org/10.3390/tomography7030038

Author

Olsson, Hampus ; Novén, Mikael ; Lätt, Jimmy ; Wirestam, Ronnie ; Helms, Gunther. / Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol. I: Tomography. 2021 ; Bind 7, Nr. 3. s. 434-451.

Bibtex

@article{a859d0310eaa40e09daa1d06a1a181a9,
title = "Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol",
abstract = "At field strengths of 7 T and above, T1-weighted imaging of human brain suffers increas-ingly from radiofrequency (RF) B1 inhomogeneities. The well-known MP2RAGE (magnetization prepared two rapid acquisition gradient echoes) sequence provides a solution but may not be readily available for all MR systems. Here, we describe the implementation and evaluation of a sequential protocol to obtain normalized magnetization prepared rapid gradient echo (MPRAGE) images at 0.7, 0.8, or 0.9-mm isotropic spatial resolution. Optimization focused on the reference gradient-recalled echo (GRE) that was used for normalization of the MPRAGE. A good compromise between white-gray matter contrast and the signal-to-noise ratio (SNR) was reached at a flip angle of 3◦ and total scan time was reduced by increasing the reference voxel size by a factor of 8 relative to the MPRAGE resolution. The average intra-subject coefficient-of-variation (CV) in segmented white matter (WM) was 7.9 ± 3.3% after normalization, compared to 20 ± 8.4% before. The corresponding inter-subject average CV in WM was 7.6 ± 7.6% and 13 ± 7.8%. Maps of T1 derived from forward signal modelling showed no obvious bias after correction by a separately acquired flip angle map. To conclude, a non-interleaved acquisition for normalization of MPRAGE offers a simple alternative to MP2RAGE to obtain semi-quantitative purely T1-weighted images. These images can be converted to T1 maps, analogously to the established MP2RAGE approach. Scan time can be reduced by increasing the reference voxel size which has only a miniscule effect on image quality.",
keywords = "7T, B1, Bias correction, Intensity correction, Longitudinal relaxation, MP2RAGE, MPRAGE, T1, T1-mapping, Ultra-high field",
author = "Hampus Olsson and Mikael Nov{\'e}n and Jimmy L{\"a}tt and Ronnie Wirestam and Gunther Helms",
note = "Publisher Copyright: {\textcopyright} 2021 by the authors. Licensee MDPI, Basel, Switzerland.",
year = "2021",
doi = "10.3390/tomography7030038",
language = "English",
volume = "7",
pages = "434--451",
journal = "Tomography - A Journal for Imaging Research",
issn = "2379-1381",
publisher = "Grapho Publications, LLC (Ann Abor, Michigan)",
number = "3",

}

RIS

TY - JOUR

T1 - Radiofrequency bias correction of magnetization prepared rapid gradient echo mri at 7.0 tesla using an external reference in a sequential protocol

AU - Olsson, Hampus

AU - Novén, Mikael

AU - Lätt, Jimmy

AU - Wirestam, Ronnie

AU - Helms, Gunther

N1 - Publisher Copyright: © 2021 by the authors. Licensee MDPI, Basel, Switzerland.

PY - 2021

Y1 - 2021

N2 - At field strengths of 7 T and above, T1-weighted imaging of human brain suffers increas-ingly from radiofrequency (RF) B1 inhomogeneities. The well-known MP2RAGE (magnetization prepared two rapid acquisition gradient echoes) sequence provides a solution but may not be readily available for all MR systems. Here, we describe the implementation and evaluation of a sequential protocol to obtain normalized magnetization prepared rapid gradient echo (MPRAGE) images at 0.7, 0.8, or 0.9-mm isotropic spatial resolution. Optimization focused on the reference gradient-recalled echo (GRE) that was used for normalization of the MPRAGE. A good compromise between white-gray matter contrast and the signal-to-noise ratio (SNR) was reached at a flip angle of 3◦ and total scan time was reduced by increasing the reference voxel size by a factor of 8 relative to the MPRAGE resolution. The average intra-subject coefficient-of-variation (CV) in segmented white matter (WM) was 7.9 ± 3.3% after normalization, compared to 20 ± 8.4% before. The corresponding inter-subject average CV in WM was 7.6 ± 7.6% and 13 ± 7.8%. Maps of T1 derived from forward signal modelling showed no obvious bias after correction by a separately acquired flip angle map. To conclude, a non-interleaved acquisition for normalization of MPRAGE offers a simple alternative to MP2RAGE to obtain semi-quantitative purely T1-weighted images. These images can be converted to T1 maps, analogously to the established MP2RAGE approach. Scan time can be reduced by increasing the reference voxel size which has only a miniscule effect on image quality.

AB - At field strengths of 7 T and above, T1-weighted imaging of human brain suffers increas-ingly from radiofrequency (RF) B1 inhomogeneities. The well-known MP2RAGE (magnetization prepared two rapid acquisition gradient echoes) sequence provides a solution but may not be readily available for all MR systems. Here, we describe the implementation and evaluation of a sequential protocol to obtain normalized magnetization prepared rapid gradient echo (MPRAGE) images at 0.7, 0.8, or 0.9-mm isotropic spatial resolution. Optimization focused on the reference gradient-recalled echo (GRE) that was used for normalization of the MPRAGE. A good compromise between white-gray matter contrast and the signal-to-noise ratio (SNR) was reached at a flip angle of 3◦ and total scan time was reduced by increasing the reference voxel size by a factor of 8 relative to the MPRAGE resolution. The average intra-subject coefficient-of-variation (CV) in segmented white matter (WM) was 7.9 ± 3.3% after normalization, compared to 20 ± 8.4% before. The corresponding inter-subject average CV in WM was 7.6 ± 7.6% and 13 ± 7.8%. Maps of T1 derived from forward signal modelling showed no obvious bias after correction by a separately acquired flip angle map. To conclude, a non-interleaved acquisition for normalization of MPRAGE offers a simple alternative to MP2RAGE to obtain semi-quantitative purely T1-weighted images. These images can be converted to T1 maps, analogously to the established MP2RAGE approach. Scan time can be reduced by increasing the reference voxel size which has only a miniscule effect on image quality.

KW - 7T

KW - B1

KW - Bias correction

KW - Intensity correction

KW - Longitudinal relaxation

KW - MP2RAGE

KW - MPRAGE

KW - T1

KW - T1-mapping

KW - Ultra-high field