Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

Standard

Overexpansion of older generation balloon expandable transcatheter heart valves : An ex-vivo bench study. / Sathananthan, Janarthanan; Sellers, Stephanie; Barlow, Aaron; Hensey, Mark; Landes, Uri; Cheung, Anson; Ye, Jian; Alkhodair, Abdullah; Blanke, Philipp; Søndergaard, Lars; Pibarot, Phillippe; Wood, David; Leipsic, Jonathan; Webb, John.

I: Catheterization and Cardiovascular Interventions, Bind 94, Nr. 6, 2019, s. 806-811.

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

Harvard

Sathananthan, J, Sellers, S, Barlow, A, Hensey, M, Landes, U, Cheung, A, Ye, J, Alkhodair, A, Blanke, P, Søndergaard, L, Pibarot, P, Wood, D, Leipsic, J & Webb, J 2019, 'Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study', Catheterization and Cardiovascular Interventions, bind 94, nr. 6, s. 806-811. https://doi.org/10.1002/ccd.28160

APA

Sathananthan, J., Sellers, S., Barlow, A., Hensey, M., Landes, U., Cheung, A., Ye, J., Alkhodair, A., Blanke, P., Søndergaard, L., Pibarot, P., Wood, D., Leipsic, J., & Webb, J. (2019). Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study. Catheterization and Cardiovascular Interventions, 94(6), 806-811. https://doi.org/10.1002/ccd.28160

Vancouver

Sathananthan J, Sellers S, Barlow A, Hensey M, Landes U, Cheung A o.a. Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study. Catheterization and Cardiovascular Interventions. 2019;94(6):806-811. https://doi.org/10.1002/ccd.28160

Author

Sathananthan, Janarthanan ; Sellers, Stephanie ; Barlow, Aaron ; Hensey, Mark ; Landes, Uri ; Cheung, Anson ; Ye, Jian ; Alkhodair, Abdullah ; Blanke, Philipp ; Søndergaard, Lars ; Pibarot, Phillippe ; Wood, David ; Leipsic, Jonathan ; Webb, John. / Overexpansion of older generation balloon expandable transcatheter heart valves : An ex-vivo bench study. I: Catheterization and Cardiovascular Interventions. 2019 ; Bind 94, Nr. 6. s. 806-811.

Bibtex

@article{dd2b292e6aa744cca281eb60c6f89f99,
title = "Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study",
abstract = "BACKGROUND: The SAPIEN 3 (S3) transcatheter heart valve (THV) can be over-expanded beyond its labeled diameter. Overexpansion can be achieved with use of either a compliant or non-compliant balloon. Objective data regarding the ability to over-expand older generation balloon expandable valves are limited. We sought to assess the effects of over-expanding the SAPIEN and SAPIEN XT (Edwards Lifesciences, Irvine, CA) valve beyond labeled size (diameter) through an ex-vivo bench study.METHODS AND RESULTS: We assessed SAPIEN and SAPIEN XT THVs, sized 23/26 mm and 23/26/29 mm, respectively. The SAPIEN THVs were explanted samples. Valves were expanded to nominal dimensions, and then incrementally overexpanded with balloons sized 1-, 2-, and 3-mm larger than the recommended diameter. When an appropriate sized non-compliant balloon was not available, a compliant balloon was utilized. Valves underwent visual and radiographic assessment of overexpansion. SAPIEN THVs with labeled size of 23 and 26 mm could be incrementally overexpanded to midvalve (MV) diameters of 26.7 and 27.4 mm, respectively. SAPIENT XT THVs with labeled size of 23, 26, and 29 mm could be incrementally overexpanded to MV diameters of 26.8, 28.3, and 28.8 mm, respectively. The desired degree of overexpansion was only achieved with use of non-compliant balloons and not with compliant balloons. The outflow of the SAPIEN and SAPIEN XT had larger diameters than the MV and inflow of the THV.CONCLUSION: Overexpansion of older generation SAPIEN and SAPIEN XT THVs is possible. Achieving the desired degree of overexpansion was only achieved with use of non-compliant balloons. This has potential implications for the treatment of failed THVs.",
author = "Janarthanan Sathananthan and Stephanie Sellers and Aaron Barlow and Mark Hensey and Uri Landes and Anson Cheung and Jian Ye and Abdullah Alkhodair and Philipp Blanke and Lars S{\o}ndergaard and Phillippe Pibarot and David Wood and Jonathan Leipsic and John Webb",
year = "2019",
doi = "10.1002/ccd.28160",
language = "English",
volume = "94",
pages = "806--811",
journal = "Catheterization and Cardiovascular Interventions",
issn = "1522-1946",
publisher = "JohnWiley & Sons, Inc.",
number = "6",

}

RIS

TY - JOUR

T1 - Overexpansion of older generation balloon expandable transcatheter heart valves

T2 - An ex-vivo bench study

AU - Sathananthan, Janarthanan

AU - Sellers, Stephanie

AU - Barlow, Aaron

AU - Hensey, Mark

AU - Landes, Uri

AU - Cheung, Anson

AU - Ye, Jian

AU - Alkhodair, Abdullah

AU - Blanke, Philipp

AU - Søndergaard, Lars

AU - Pibarot, Phillippe

AU - Wood, David

AU - Leipsic, Jonathan

AU - Webb, John

PY - 2019

Y1 - 2019

N2 - BACKGROUND: The SAPIEN 3 (S3) transcatheter heart valve (THV) can be over-expanded beyond its labeled diameter. Overexpansion can be achieved with use of either a compliant or non-compliant balloon. Objective data regarding the ability to over-expand older generation balloon expandable valves are limited. We sought to assess the effects of over-expanding the SAPIEN and SAPIEN XT (Edwards Lifesciences, Irvine, CA) valve beyond labeled size (diameter) through an ex-vivo bench study.METHODS AND RESULTS: We assessed SAPIEN and SAPIEN XT THVs, sized 23/26 mm and 23/26/29 mm, respectively. The SAPIEN THVs were explanted samples. Valves were expanded to nominal dimensions, and then incrementally overexpanded with balloons sized 1-, 2-, and 3-mm larger than the recommended diameter. When an appropriate sized non-compliant balloon was not available, a compliant balloon was utilized. Valves underwent visual and radiographic assessment of overexpansion. SAPIEN THVs with labeled size of 23 and 26 mm could be incrementally overexpanded to midvalve (MV) diameters of 26.7 and 27.4 mm, respectively. SAPIENT XT THVs with labeled size of 23, 26, and 29 mm could be incrementally overexpanded to MV diameters of 26.8, 28.3, and 28.8 mm, respectively. The desired degree of overexpansion was only achieved with use of non-compliant balloons and not with compliant balloons. The outflow of the SAPIEN and SAPIEN XT had larger diameters than the MV and inflow of the THV.CONCLUSION: Overexpansion of older generation SAPIEN and SAPIEN XT THVs is possible. Achieving the desired degree of overexpansion was only achieved with use of non-compliant balloons. This has potential implications for the treatment of failed THVs.

AB - BACKGROUND: The SAPIEN 3 (S3) transcatheter heart valve (THV) can be over-expanded beyond its labeled diameter. Overexpansion can be achieved with use of either a compliant or non-compliant balloon. Objective data regarding the ability to over-expand older generation balloon expandable valves are limited. We sought to assess the effects of over-expanding the SAPIEN and SAPIEN XT (Edwards Lifesciences, Irvine, CA) valve beyond labeled size (diameter) through an ex-vivo bench study.METHODS AND RESULTS: We assessed SAPIEN and SAPIEN XT THVs, sized 23/26 mm and 23/26/29 mm, respectively. The SAPIEN THVs were explanted samples. Valves were expanded to nominal dimensions, and then incrementally overexpanded with balloons sized 1-, 2-, and 3-mm larger than the recommended diameter. When an appropriate sized non-compliant balloon was not available, a compliant balloon was utilized. Valves underwent visual and radiographic assessment of overexpansion. SAPIEN THVs with labeled size of 23 and 26 mm could be incrementally overexpanded to midvalve (MV) diameters of 26.7 and 27.4 mm, respectively. SAPIENT XT THVs with labeled size of 23, 26, and 29 mm could be incrementally overexpanded to MV diameters of 26.8, 28.3, and 28.8 mm, respectively. The desired degree of overexpansion was only achieved with use of non-compliant balloons and not with compliant balloons. The outflow of the SAPIEN and SAPIEN XT had larger diameters than the MV and inflow of the THV.CONCLUSION: Overexpansion of older generation SAPIEN and SAPIEN XT THVs is possible. Achieving the desired degree of overexpansion was only achieved with use of non-compliant balloons. This has potential implications for the treatment of failed THVs.

U2 - 10.1002/ccd.28160

DO - 10.1002/ccd.28160

M3 - Journal article

C2 - 30851079

VL - 94

SP - 806

EP - 811

JO - Catheterization and Cardiovascular Interventions

JF - Catheterization and Cardiovascular Interventions

SN - 1522-1946

IS - 6

ER -

ID: 241106097