Intra- and inter-observer agreement and reliability of bone mineral density measurements around acetabular cup: a porcine ex-vivo study using single- and dual-energy computed tomography
Publikation: Bidrag til tidsskrift › Tidsskriftartikel › Forskning › fagfællebedømt
Standard
Intra- and inter-observer agreement and reliability of bone mineral density measurements around acetabular cup : a porcine ex-vivo study using single- and dual-energy computed tomography. / Mussmann, Bo Redder; Overgaard, Soren; Torfing, Trine; Bøgehøj, Morten Foged; Gerke, Oke; Andersen, Poul Erik.
I: Acta Radiologica Open, Bind 6, Nr. 7, 2017, s. 1-9.Publikation: Bidrag til tidsskrift › Tidsskriftartikel › Forskning › fagfællebedømt
Harvard
APA
Vancouver
Author
Bibtex
}
RIS
TY - JOUR
T1 - Intra- and inter-observer agreement and reliability of bone mineral density measurements around acetabular cup
T2 - a porcine ex-vivo study using single- and dual-energy computed tomography
AU - Mussmann, Bo Redder
AU - Overgaard, Soren
AU - Torfing, Trine
AU - Bøgehøj, Morten Foged
AU - Gerke, Oke
AU - Andersen, Poul Erik
PY - 2017
Y1 - 2017
N2 - Background: Periprosthetic bone loss is considered to be a potentially contributing factor in aseptic loosening of acetabular hip components, but no studies have shown this association. The lack of association might be caused by insufficient image quality because of metal artifacts and challenges in measuring bone density (BMD) in complex anatomic structures which might be overcome using dual-energy computed tomography (DECT).PurposeTo test inter- and intra-observer agreement and reliability of in-house segmentation software measuring BMD adjacent to acetabular cup and to compare measurements performed with single-energy CT (SECT) and DECT in cemented and cementless cups.Material and Methods: Twenty-four acetabular cups inserted in porcine hip specimens were scanned with SECT and DECT. Bone density was measured in a three-dimensional volume adjacent to the cup. Double measurements were performed.Results: BMD derived from SECT was approximately four times higher than that of DECT. In both scan modes, intraclass correlation coefficient (ICC) was >0.90 with no differences between repeated measurements, except for uncemented cups where a statistically significant difference of 11 mg/cm3 was found with DECT. DECT showed narrower limits of agreement than SECT. Inter-observer analysis showed small differences.Conclusion: BMD can be estimated with high intra- and inter-observer reliability with SECT and DECT around acetabular cups using custom software. The intra- and inter-observer agreement of DECT is superior to that of SECT and better in the cementless concept. Good intra- and inter-observer reliability can be obtained in both cemented and cementless cups using the segmentation software. SECT and DECT cannot be used interchangeably.
AB - Background: Periprosthetic bone loss is considered to be a potentially contributing factor in aseptic loosening of acetabular hip components, but no studies have shown this association. The lack of association might be caused by insufficient image quality because of metal artifacts and challenges in measuring bone density (BMD) in complex anatomic structures which might be overcome using dual-energy computed tomography (DECT).PurposeTo test inter- and intra-observer agreement and reliability of in-house segmentation software measuring BMD adjacent to acetabular cup and to compare measurements performed with single-energy CT (SECT) and DECT in cemented and cementless cups.Material and Methods: Twenty-four acetabular cups inserted in porcine hip specimens were scanned with SECT and DECT. Bone density was measured in a three-dimensional volume adjacent to the cup. Double measurements were performed.Results: BMD derived from SECT was approximately four times higher than that of DECT. In both scan modes, intraclass correlation coefficient (ICC) was >0.90 with no differences between repeated measurements, except for uncemented cups where a statistically significant difference of 11 mg/cm3 was found with DECT. DECT showed narrower limits of agreement than SECT. Inter-observer analysis showed small differences.Conclusion: BMD can be estimated with high intra- and inter-observer reliability with SECT and DECT around acetabular cups using custom software. The intra- and inter-observer agreement of DECT is superior to that of SECT and better in the cementless concept. Good intra- and inter-observer reliability can be obtained in both cemented and cementless cups using the segmentation software. SECT and DECT cannot be used interchangeably.
UR - http://journals.sagepub.com/doi/full/10.1177/2058460117719746
U2 - 10.1177/2058460117719746
DO - 10.1177/2058460117719746
M3 - Journal article
C2 - 28785480