A comparison between two NIRS oximeters (INVOS, OxyPrem) using measurement on the arm of adults and head of infants after caesarean section
Research output: Contribution to journal › Journal article › Research › peer-review
Standard
A comparison between two NIRS oximeters (INVOS, OxyPrem) using measurement on the arm of adults and head of infants after caesarean section. / Hyttel-Sørensen, Simon; Hessel, Trine Witzner; la Cour, Amalia; Greisen, Gorm.
In: Biomedical Optics Express, Vol. 5, No. 10, 2014, p. 3671-3683.Research output: Contribution to journal › Journal article › Research › peer-review
Harvard
APA
Vancouver
Author
Bibtex
}
RIS
TY - JOUR
T1 - A comparison between two NIRS oximeters (INVOS, OxyPrem) using measurement on the arm of adults and head of infants after caesarean section
AU - Hyttel-Sørensen, Simon
AU - Hessel, Trine Witzner
AU - la Cour, Amalia
AU - Greisen, Gorm
PY - 2014
Y1 - 2014
N2 - Previously the NIRS oximeter OxyPrem was calibrated by comparison to the INVOS in a blood-lipid phantom. The aim of the present study was to test this calibration clinically. During vasculur occlusions in 10 adults and after birth in 25 term infants the relationship was OxyPrem = 1.24 x INVOS - 23.6% and OxyPrem = 1.15 x INVOS - 16.2% on the adult arm and infant head, respectively. The precsion during steady state was 4.0% (CI 3.4% to 4.6%) and 3.4% (CI 2.9% to 3.9%) on the arm, and 6.7% (CI 5.9% to 7.6%) and 4.7% (CI 3.5% to 5.9%) on the infant head for OxyPrem and INVOS, respectively. We conclude that the calibration on the blood-lipid phantom was unsuccessful in achieving agreement in clinical measurements.
AB - Previously the NIRS oximeter OxyPrem was calibrated by comparison to the INVOS in a blood-lipid phantom. The aim of the present study was to test this calibration clinically. During vasculur occlusions in 10 adults and after birth in 25 term infants the relationship was OxyPrem = 1.24 x INVOS - 23.6% and OxyPrem = 1.15 x INVOS - 16.2% on the adult arm and infant head, respectively. The precsion during steady state was 4.0% (CI 3.4% to 4.6%) and 3.4% (CI 2.9% to 3.9%) on the arm, and 6.7% (CI 5.9% to 7.6%) and 4.7% (CI 3.5% to 5.9%) on the infant head for OxyPrem and INVOS, respectively. We conclude that the calibration on the blood-lipid phantom was unsuccessful in achieving agreement in clinical measurements.
U2 - 10.1364/BOE.5.003671
DO - 10.1364/BOE.5.003671
M3 - Journal article
C2 - 25360381
VL - 5
SP - 3671
EP - 3683
JO - Biomedical Optics Express
JF - Biomedical Optics Express
SN - 2156-7085
IS - 10
ER -
ID: 137497226