Automatic correction of dental artifacts in PET/MRI
Publikation: Bidrag til tidsskrift › Tidsskriftartikel › Forskning › fagfællebedømt
Dokumenter
- Automatic correction of dental artifacts in PET/MRI
Forlagets udgivne version, 5,7 MB, PDF-dokument
A challenge when using current magnetic resonance (MR)-based attenuation correction in positron emission tomography/MR imaging (PET/MRI) is that the MRIs can have a signal void around the dental fillings that is segmented as artificial air-regions in the attenuation map. For artifacts connected to the background, we propose an extension to an existing active contour algorithm to delineate the outer contour using the non-attenuation corrected PET image and the original attenuation map. We propose a combination of two different methods for differentiating the artifacts within the body from the anatomical air-regions by first using a template of artifact regions, and second, representing the artifact regions with a combination of active shape models and k-nearest-neighbors. The accuracy of the combined method has been evaluated using 25 18 F-fluorodeoxyglucose PET/MR patients. Results showed that the approach was able to correct an average of 97+/- 3% of the artifact areas.
Originalsprog | Engelsk |
---|---|
Artikelnummer | 024009 |
Tidsskrift | SPIE Journal of Medical Imaging |
Vol/bind | 2 |
Udgave nummer | 2 |
Antal sider | 13 |
ISSN | 2329-4302 |
DOI | |
Status | Udgivet - 2015 |