Online monitoring of urea concentration in dialysate with dual-beam Fourier-transform near-infrared spectroscopy

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningfagfællebedømt

The robustness of a dual-beam, optical null, Fouriertransform near-infrared (FTNIR) spectrometer was investigated by, means of online, near-infrared measurements and predictions of urea concentrations in spent dialysate during hemodialysis treatment. Simple multivariate calibration using a few factors based on a small number of prepared samples provided stable and accurate predictions over a period of 1 month. The calibration was robust when faced with adjustment of reference cell intensity and did not require a daily measured reference spectrum. The root-mean-square error of prediction of urea was 0.4 mM based on a two-factor partial least-squares regression model.

OriginalsprogEngelsk
TidsskriftJournal of Biomedical Optics
Vol/bind9
Udgave nummer3
Sider (fra-til)553-557
Antal sider5
ISSN1083-3668
DOI
StatusUdgivet - maj 2004

Bibliografisk note

Funding Information:
The authors gratefully acknowledge the participation of patients and personnel at the Copenhagen University Hospital and thank Foss Electric A/S for access to their thermostatted flow cell for this experiment. They further gratefully acknowledge financial support from the Danish Pharmaceutical Association, the Augustinus Foundation, and the Director Ib Hen-riksen Foundation.